GE Analytical Instruments ©2016 153 of 270 DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
how to import this data into a spreadsheet program, see "Appendix C: Importing Analyzer Data into a
Spreadsheet."
To export real-time data, refer to “Collecting Historical and Real-Time Data” on page 159.
Figure 59: Export (Data) Screen
To export data history
To export data using the Ethernet port, you must first configure the Analyzer's IP address (see “Connecting the
Analyzer to a Network Via Ethernet” on page 156).
1. Select the Data tab, and then press the Export button. The Export screen appears.
2. Press the Range button to select the time frame for the data to export. The following buttons appear on
the right:
•1 Day
• 1 Week
•All
• Custom (Enter a specific date and time range.)