GE Analytical Instruments ©2016 121 of 270 DLM 68100-09 EN Rev. A
Chapter 5: Calibration and Verification
Figure 37: Editing a Calibration
Performing a Calibration Verification 
Verification parallels the options available under calibration menu choices. Verifications may be performed in 
TOC, NPOC, TC, or IC analysis modes. As with calibration, you can save multiple verification protocols in each of 
the Analyzer’s five ranges, and you can save up to 20 verification protocols in total.
For more detailed information regarding options available under verification menus, consult the relevant 
sections describing calibration, keeping in mind that you must press the Verification button rather than the 
Calibrate button in the Cal/Ver tab, and for each verification protocol you must specify a calibration protocol for 
the measurements.
Table 10
 lists the approximate number of repetitions that are possible from a 40-mL vial, completely filled with 
the standard solution. If Auto Dilution is employed, more repetitions are possible than are shown in Table 10
, but 
the maximum number of repetitions that can be selected is 20.