GE Analytical Instruments ©2016 127 of 270 DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
Continuous Measurements
By turning Continuous On, the Analyzer makes continuous measurements of the enabled sample streams and 
the Check Standard (if it is enabled). Continuous measurement of Stream 1 is shown in Figure 41
. Only Stream 1 
and the Check Standard are enabled in this example, so measurements of Stream 1 are made without a pause, 
with two exceptions.
Figure 41: Continuous Operation
The first exception is the Check Standard, since it is enabled. Measurements of the Check Standard are made 
according to the schedule established by the operator. In the example shown in Figure 42
, the check standard is 
measured every three hours (the Check Standard Frequency). Since two Repetitions have been selected, the 
check standard is measured twice every three hours. The first measurement is ignored because it has been 
designated as a Reject.
Figure 42: Operation with a Cycle Frequency of One Hour