GE Analytical Instruments ©2016 185 of 270 DLM 68100-09 EN Rev. A
Chapter 8: Maintenance
5. Using a 3/32 Allen wrench, loosen and remove the screw that secures the syringe. Remove any washers
under the block, and retain them for re-installation
Figure 75: Locating the Syringe Screw and Washers
6. Unscrew the syringe base by turning the base to the left.
Figure 76: Unscrewing the Syringe Base
Washers*
*If these washers are NOT installed
with your Analyzer. ignore this step.