GE Analytical Instruments ©2016 155 of 270 DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
Reviewing Exported Data
When the Analyzer exports data via the USB or Ethernet ports, the export data includes the following:
Table 13: Data Fields Output
Field Value Output Options (if Available)
1 Date MM/DD/YYYY HH:MM
2 Sample Name Stream 1-5, Check Standard, User-Entered Sample Name (Up to
30 characters)
3 Analysis Mode NPOC, TOC, IC, TC
4 Stream Number Blank, 1, 2, 3, 4, or 5
5NPOC (ppm)
6TOC (ppm)
7 IC (ppm)
8 TC (ppm)
9 Converted Result
10 Unit
11 Reject Y or N
12 Acid (%)
13 Oxid (%)
14 Range Mode Manual or Auto
15 Range Up to 100, Up to 1,000, Up to 5,000, Up to 20,000, Up to 50,000
16 Sparge (min.)
17 Flush OFF, DI, 1S, 2S, 3S, 4S, or 5S
18 Blank Correction ON or OFF
19 Resuspend OFF
20 Pressure (psi) Ambient pressure, psia
21 Man Dilution Manual Dilution Numeric Value
22 Calibration Name Alphanumeric characters
23 Calibration Date MM/DD/YYYY HH:MM
24 Error/Warning Error/Warning Number
25 CO
2
(mg) Mass of CO
2
measured (mg). Rejects are not reported.
26 Mass Response (µg)
27 Equation Conversion factor equation
28 Method ID Conversion factor name
29 Leak Check Ratio Diagnostic should be <1.008