Chapter 6: Analyzer Operation
Figure 62: Alarm Output Screen
3. Press the Alarm Stream button. Buttons for each stream appear to the right.
4. Press the stream button you want to the alarm to respond to (such as the Stream 1 button or Stream 2
button).
5. Press the High Level Alarm button and use the numeric keypad that appears to the right to set the
threshold concentration (in ppm units
10
). Any measurement above this threshold will trigger the alarm.
6. Press the Low Level Alarm button and use the numeric keypad that appears to the right to set the
threshold concentration (in ppm units
10.
). Any measurement below this threshold will trigger the alarm.
7. To activate the alarms, press the Alarm On/Off button, and then press the On button that appears.
10.
The unit of measure appears in the default ppm unit or another unit measure if the stream’s Setup screen
has been configured with a conversion factor.
NOTE:
The low-level set point value must be less than the high-level set point value.
NOTE:
The high and low threshold concentrations refer to the parameter the stream is configured to
measure. For example, if Alarm 1 responds to Stream 1, and Stream 1 is configured to measure
TOC, then the High Level Alarm and the Low Level Alarm are in the units of ppm TOC.