GE Analytical Instruments ©2016 74 of 270 DLM 68100-09 EN Rev. A
Chapter 4: Installation
Figure 2: Mounting Stand - Clearances
At least two people are required to move the Mounting Stand to the selected location. Bolt the Mounting Stand
to a wall or the floor before placing the Analyzer on it. The horizontal braces shown in Figure 2
can be adjusted
so that the display of the Analyzer is approximately eye-level, but the spacing between the two horizontal
braces must be 94.5 cm.
To mount a Filter Panel on the Mounting Stand, attach the two Filter Panel brackets 30.5 cm apart, as shown in
Figure 2
. Further instructions for installation of the Filter Panel are provided in the GE InnovOx Filter Panel
Operation and Maintenance Manual.
12.0 in. (30.5 cm.)
37.21 in. (94.5 cm.)
28.38 in. (72.1 cm.)
35.5 in. (90.2 cm.)
71.3 in. (181.3 cm.)
Mounting brackets for
filter panel (optional)
44.1 in. (112.0 cm.)