GE Analytical Instruments ©2016 147 of 270 DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
Figure 52: The Maintenance Tab
1. When the User Management feature is enabled, first log in to the InnovOx with the Administrator ID and
password. Otherwise, if the User Management feature is disabled, continue to the next step.
2. Select the Maintenance tab and press the Conversion Factor button (Figure 52
) to access the five
available conversion factors (Factor 1, Factor 2, Factor 3, Factor 4, and Factor 5).
3. Press the Factor 1 button (or other Factor button to configure that factor). The Factor screen appears
with configurable Factor Equation, Factor Units and Factor Name buttons (Figure 53
).
Figure 53: Configuring a Conversion Factor
4. To set up a Conversion Factor equation, press the Factor Equation button and use the keypad to enter
the equation. Use the letter "C" in the equation to represent the raw value produced by the Analyzer. (It