GE Analytical Instruments ©2016 131 of 270 DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
Online Measurement—Configuration
Figure 45: Configure Online Screen
You can use the Configure Online screen (Setup tab Online button Configure Online) to set up online
measurements for the following:
Operation with Multiple Streams
Check Standards (Scheduled or Manual)
Continuous (online measuring)
Cycle Frequency
Clean Inlet
•Auto Restart
Daily Flush
Daily Flush Start
You can also calculate Run Time and view a Summary of online configuration on this screen.
Operation with Multiple Streams
The InnovOx On-Line TOC Analyzer may be used to monitor multiple streams. The base configuration can
monitor two streams, and five streams can be monitored with the optional 5-Stream configuration. With the
exception of the number of streams, both Analyzers are set up the same way.
NOTE:
The Fail Safe button, previously located on this screen, has been moved to the
Advanced screen (Maintenance tab > Advanced). For information on this feature, refer
to
Setting the Fail Safe Option” on page 169.