EasyManua.ls Logo

GE Sievers Innovox - Calibration and Verification Options

GE Sievers Innovox
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2016 107 of 270 DLM 68100-09 EN Rev. A
Chapter 5: Calibration and Verification
For accuracy, the IC standard solution(s) must be prepared just before the calibration. The solutions are basic, so
it will absorb CO
2
from the air. That absorption can significantly change the concentration of the solution. To
minimize the rate of CO
2
absorption, the solution should be stored in a sealed container until use. Just prior to
the IC calibration, a 40-mL vial should be filled with the standard solution.
Calibration and Verification Options
When calibrating the InnovOx On-Line Analyzer using either the Single-Point Calibration or Two-Point
Calibration, the algorithm used to calculate the slope of the calibration curve uses the following variables:
The mass response
The ratio of sample volume to total volume (where total volume includes sample, dilution water,
reagents, and so on.)
The concentration of the standard (e.g., 1000 ppm)
Slope
7
= [(standard concentration) * (V
s
/V
t)
]/(mass response)
Single-Point Calibration
When a single-point calibration is performed, only one standard is used, as shown in Figure 31. For that reason,
single-point calibrations are convenient and rapid; however, the analysis of samples may be less accurate than
when other calibration methods are used.
7.
V
S
= Volume of sample in sparger needed for dilution range, mL
V
T
= Total solution volume in sparger, mL

Table of Contents

Related product manuals