EasyManua.ls Logo

GE Sievers Innovox - Point-To-Point Multi-Point Calibration

GE Sievers Innovox
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2016 112 of 270 DLM 68100-09 EN Rev. A
Chapter 5: Calibration and Verification
Figure 34: Linear Fit Six-Point Calibration
Point-to-Point Multi-Point Calibration
The advantage of selecting the point-to-point calibration is that any non-linearity of response is compensated
for during the calibration.
Figure 35
depicts a point-to-point calibration with 6 points. Like the linear fit, the point having the lowest
concentration is measured first, and the points must increase in concentration at each step. Point 1 can either
be reagent water or a higher, known concentration.
The Analyzer calculates an intercept and slope for the line connecting each pair of points, as shown in Figure 35
.
When this calibration is later applied to a sample, the Analyzer applies the appropriate intercept and slope,
based on the magnitude of its mass response for that sample.

Table of Contents

Related product manuals