GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 3 - Setup Instructions 3-51
3-8-1-5 How to activate Screen Lock
1.) Invoke System Setup as described in Section 3-8-1-1 on page 3-49.
2.) On the left side of the screen select GENERAL
and then click on the GENERAL tab.
3.) Check mark “Screen Lock”.
a.) If no password previously entered, the following dialog appears.
b.) Enter “New Password”.
NOTE: A new screen lock password must be at least 6 characters long and has a maximum length of 80
characters. The password must contain at least 2 non-letter characters, 0...9 or !@#$%^*().
c.) “Retype new Password” and then click SAVE&EXIT
to save new screen lock password.
4.) Click SAVE&EXIT
to save Settings and exit System Setup.
NOTE: If screen is locked you have to enter the password to get full system control.
If password is unknown click EMERGENCY
. This enables standard - but limited - operation.
Figure 3-52 Screenlock Password
NOTICE
!! NOTICE:
The Screen Lock password cannot be reset by the user! Please contact your GE service representative.