GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 2 - Site Preparation 2-9
2-3-4 Networking Setup Requirements
2-3-4-1 Stand-alone system (without Network Connection)
None
2-3-4-2 Unit Connected to Hospital’s Network
Supported networks:
Ethernet
Wireless LAN
2-3-4-3 Purpose of the DICOM Network Function
DICOM (Digital Imaging and Communications in Medicine) services provide the operator with clinically
useful features for moving images and patient information over a hospital network.
Examples of DICOM services include the transfer of images to workstations for viewing or transferring
images to remote printers. As an added benefit, transferring images in this manner frees up the on-
board monitor and peripherals, enabling viewing to be done while scanning continues.
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost.
2-3-4-4 DICOM Option Pre-installation Requirements
To configure the Voluson E8 / Voluson E6 ultrasound system to work with other network connections,
the network administrator must provide some necessary information.
Use the Connectivity Setup Worksheet on page 3-68 to record required information that must include:
Installation see: Section 3-12 "Network IP Address Configuration" on page 3-65.
• Voluson E8 / Voluson E6
Details:
DICOM network details for the Voluson system, including the
host name, local port, IP address, AE title and net mask.
• Routing Information: IP addresses for default gateway and other routers in use at site.
• DICOM Application Information: Details of DICOM devices in use at the site, including the DICOM
host name, AE title, DICOM port number and IP addresses.