3-8 Section 3-4 - Preparing for Set Up
3-4-2 System Voltage Settings
Verify that the scanner is set to the correct voltage. The Voltage setting for the Voluson E8 / Voluson
E6 Scanner is found on the identification plate (see: Figure 1-2 on page 1-14) on the rear of the system.
3-4-3 EMI Protection
This system has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of
the covers, shields, and screws are provided primarily to protect the system from image artifacts caused
by this interference. For this reason, it is imperative that all covers and hardware are installed and
secured before the system is put into operation.
Ensure that the system is protected from electromagnetic interference (EMI), as follows:
Operate the system at least 15 feet away from equipment that emits strong electromagnetic
radiation.
Operate the system in an area enclosed by walls, floors and ceilings comprised of wood, plaster or
concrete, which help prevent EMI.
Shield the system when operating it in the vicinity of radio broadcast equipment, if necessary.
Do not operate mobile phones or other EMI emitting devices in the ultrasound room.
Verify that all EMI rules listed in the following table are followed:
The Voluson E8 / Voluson E6
ultrasound system is approved for use in hospitals, clinics and other
environmentally qualified facilities, in terms of the prevention of radio wave interference. Operation of
the ultrasound system
in an inappropriate environment can cause electronic interference to radios and
television sets situated near the medical equipment.
For further details and EMI Prevention/Abatement refer to Section 2-2-3 "EMI Limitations" on page 2-4.
WARNINGWARNING
!! WARNING:
CONNECTING A Voluson E8 / Voluson E6 SCANNER TO THE WRONG VOLTAGE
LEVEL WILL MOST LIKELY DESTROY THE SCANNER.