GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
7-22 Section 7-8 - Troubleshooting Trees, Instructions and Tech Tips
7-8-1 System does not boot up
Figure 7-25 System does not Power On / Boot Up
System doesn’t boot
Is Standby Switch
illuminated ?
Verify that
- Power Cord is connected to Mains
- Power Switch is turned on
(on the backside of the system)
Is Standby Switch
orange ?
Press Standby Switch till it
changes to green
System starts booting
Press Standby Switch
for approximately 5 sec.
until system is down.
Check cables between
Console and Monitor
Is there any picture on
the monitor ?
Is there any picture on
the touch screen ?
Shut off the power switch
Wait for 10 sec.
Shut on the power switch again
Is there any picture on
the monitor ?
Other error. Please
contact the service
engineer