GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-19
Section 8-10
Replacement of the Probe Holder (Kit)
8-10-1 Manpower
One person, 1 minute
8-10-2 Tools
None
8-10-3 Probe Holder (Kit)- Removal Procedure
1.) Pull out the elastic probe holder to be replaced.
8-10-4 Probe Holder (Kit) - Installation Procedure
1.) Insert the new probe holder from the kit.
Figure 8-20 pull out the probe holder
pull out the probe holder