8-12 Section 8-4 - Image Settings Only - Loading Procedure
Section 8-4
Image Settings Only - Loading Procedure
8-4-1 Introduction
The Image Settings contain:
Application Settings
User Programs
Auto Text
3D/4D Programs
Scan Assistant
8-4-2 Loading Procedure
see: Section 4-5-2 "Load Image Settings Only / Load Small Backup (Scan Settings)" on page 4-34
Section 8-5
Full Backup (Full System Configuration) - Loading Procedure
8-5-1 Introduction
The Full Backup contains following data:
User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text
entries, gray curves and complete System settings such as language, time/date format, etc.)
Measure Configuration (user specific measure setup settings)
Patient Archive (database containing patient demographic data and measurements) - no images
Options (Permanent Key that is specific for enabled software options and Demo Key)
Image Transfer Configuration (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
Network Configuration (network settings including the computer name)
Service Platform (state of the Service Software)
8-5-2 Loading Procedure
see: Section 4-5-4 "Load Full System Configuration (Full Backup)" on page 4-39
Section 8-6
Image Archive - Loading Procedure
8-6-1 Introduction
A backup of the Image Archive contains the Patient Archive (database containing patient demographic
data and measurements) + images.
8-6-2 Loading Procedure
see: Section 4-5-6-2 "Load Image Archive" on page 4-44