IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
4-34 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-2 Load Image Settings Only / Load Small Backup (Scan Settings)
4-5-2-1 Preparations
1.) Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2.) Press the UTILITIES
key on the control panel.
3.) Touch SYSTEM SETUP
button to invoke the setup desktop.
4.) On the left side of the screen select BACKUP
and then click the SYSTEM CONFIGURATION tab.
5.) Click the LOAD
button of the “Image Settings Only” resp. “Small Backup (Scan Settings)” group.
6.) Choose the media and click LOAD
.
NOTE: If it is desired to load settings from media Internal HDD, click on the CHANGE FOLDER
button, browse
for the folder “D:\usersettings\FactoryDefault\V830” and then click the LOAD
button.
Two Settings are available which slightly differ:
- General = for rest of the world
- USA = for United States (e.g., image orientation UP/DOWN inverted at transvaginal probes)
7.) Select the appropriate file and click OK
.
8.) Select the desired loading procedure:
- Load “Complete Backup” on page 4-35
- Load only parts of the “Complete Backup” on page 4-36
!! CAUTION:
The loading procedure overwrites existing image/scan settings on the local hard drive.
Make sure to insert the correct System DVD.
Additionally you can load the image settings from D:\usersettings\FactoryDefault\V830
Figure 4-24 Load window
Choose media
and
then click [Load]