GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 4 - Functional Checks 4-33
4-5-1 Save Image Settings Only / Save Small Backup (Scan Settings)
The Image/Scan Settings contain:
• Application Settings
• User Programs
• Auto Text
• 3D/4D Programs
• Scan Assistant
• Measure Configuration
• Biopsy Lines
1.) Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2.) Press the UTILITIES
key on the control panel.
3.) In the “Utilities” menu touch SYSTEM SETUP
to invoke the setup desktop on the screen.
4.) On the left side of the screen select BACKUP
and then click the SYSTEM CONFIGURATION tab.
5.) Click the SAVE
button of the “Image Settings Only” resp. “Small Backup (Scan Settings)” group.
6.) Choose the media and click the SAVE
button.
7.) Select the NEW FILE...
key and enter a file name (without extension).
8.) Click the OK
key to start the process. When the saving has been completed, click OK.
Figure 4-23 Save window
Choose media
and
then click [Save]