GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
4-32 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
Section 4-5
Backup and Restore Database, Preset Configurations and Images
1.) Press the UTILITIES key on the control panel.
2.) In the “Utilities” menu touch SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP
and then click the SYSTEM CONFIGURATION tab.
The
SYSTEM CONFIGURATION page is subdivided in 2 main groups:
A.) Image Settings Only respectively Small Backup (Scan Settings)
• Section 4-5-1 "Save Image Settings Only / Save Small Backup (Scan Settings)" on page 4-33
• Section 4-5-2 "Load Image Settings Only / Load Small Backup (Scan Settings)" on page 4-34
B.) Full System Configuration
• Section 4-5-3 "Save Full System Configuration (Full Backup)" on page 4-37
• Section 4-5-4 "Load Full System Configuration (Full Backup)" on page 4-39
• Section 4-5-5 "Delete Full System Configuration (Full Backup)" on page 4-41
Settings and/or Full System Configuration can be saved to the following destinations:
• D: partition of internal hard disk
• DVD/CD+R/RW
• Mapped Network Drive, see: Section 3-12-1 "Map Network Drive" on page 3-66
• Any other drive connected to the system (e.g.; USB-Stick or external hard disk drive)
Note:
This function is only available in the Full Backup utility.
For further details review: Section 3-5-13 on page 3-30
!! CAUTION:
It is recommended to Backup the Full System Configuration (Section 4-5-3 on page 4-37) and the
Image Archive (Section 4-5-6 on page 4-42) once a week.