GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-13
7-5-2-1 If GE Service Technician requests Remote Access Permission
If a GE Service technician requests remote access to your Voluson E8 / Voluson E6 scanner, following
“InSite Notification” appears on the systems s creen.
1.) Enable “Disruptive Mode” feature by confirming YES
.
NOTE: If the customer does not wish to have diagnostics running at the time of the request, they select NO
.
A message is sent back to the OLC or FE that “Disruptive Mode” is not enabled.
Figure 7-15 Insite Notification
select [Yes] to enable
“Disruptive Mode” feature