EasyManua.ls Logo

GE Voluson E6

GE Voluson E6
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-35
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for LongLineMemFIFO restart the system
CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for Mode_4D restart the system
CRI: mismatch in steering angle header info
Action: restart the system
restart the system
CtrlMvGrpRepresentations::isInWriteMode not supported for mode nr xx restart the system
CVIE parameter file could not be created restart the system
CW Archive::not implemented Action: restart the system restart the system
CW quant TX Frequ out of range restart the system
CW_BaseLinePos Out Of Range press ok and save this user-setting once again
CW_Gain Out Of Range press ok and save this user-setting once again
CW-hardware doesn't support
pencil probe + CW-Hardware not available->
restart the system
CW-TxFrequency == 0 restart the system
D:\\SERVICE\\ConfIndexFile is corrupted Action: restart the system restart the system
Data are not saved or sent to DICOM Server! restart the system
dDistanceFocusReferenceBF_ToProbeSurface out of valid range Action: restart the
system
restart the system
Delete error (Backup data could not be deleted.)
check storage destination of Full Backup
(e.g., DVD/CD, insufficient rights on target Network drive,
write protection on USB-drive)
Density Out Of Range press ok and save this user-setting once again
Device is not a DVD/CD rec. reconnect the device
Different software error (Backup data was made on another system with a different
software version.)
This backup cannot be restored by the user.
Disc is full! use a new CDR for writing data to CD
Display:Rect Region fails restart the system
Do you really want to delete?
Do you really want to format this DVD?
done is low! restart the system
Doppler-Txfrequency out of range restart the system
Downgrade error (Backup data was made with a software version higher than the
installed version.)
load appropriate backup for installed version
DVD recorder communication error! Check recorder cables and try again.
Make sure DVD recorder is connected to US machine.
check DVD connections, check if DVD is switched on
DVD recorder not properly connected or turned off!
Check connection from US system to recorder, recorder has power and is on.
check DVD connections, check if DVD is switched on
Error Messages Actions