GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
Peripheral Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Surface Probe Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Endocavity Probe
Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Measurement Accuracy
Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Probe/Phantom Checks
Check Gain and TGC changes, vary the
focus and check reaction on screen.
NOTE! The use of a Phantom is not required
during Preventive Maintenance. Customer
may use it as part of their Quality Assurance
Program tests.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks of all
probes Section 10-5-2
on page 10-7
•
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-1 Customer Care Schedule
Item Service at Indicated Time D W M A Notes