GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-11
10-5-5-3 Basic Probe Cleaning and/or Disinfection (cont’d)
10-5-5-4 Disinfection by means of the RIC-Holder
Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when
transporting the system with the probe attached, or during the disinfection process.
Inadequate handling may lead to dead elements, shocked head mechanics, etc.
The RIC-Holder (especially developed for RIC Real-time 4D endocavity probes) guarantees that the
sensitive probe head does not hit the bottom of the container during the disinfection procedure.
NOTE: Instructions are supplied with each RIC-Holder (KTZ225469).
CAUTION
CAUTION:
Please be aware of the
sensitive probe head.
TAKE EXTREME CARE!
NEVER place or store a
probe on its scan head!
When
disinfecting a probe, ensure
that there is sufficient space
between the probe and the
OKOK OKOK
RIC-Holder
Endocavity soaking cup kit
RIC-Holder
RIC-Holder