GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
10-20 Section 10-7 - Electrical Safety Tests
Table 10-17 shows a typical format for recording the patient lead to ground leakage current.
Measurements should be recorded from each lead and lead combination under each set of test
conditions specified in:
Table 10-10 on page 10-13, or:
Table 10-11 on page 10-13, as applicable.
Record all data on the Inspection Certificate.
Table 10-17 Typical Data Format for Recording Patient Lead to Ground Leakage
System Under Test: ________________________ Date of Test: _____________________
NOTE: Values in italics font are given as examples only.
ON NORM OPEN
ON REVERSE CLOSED
ON REVERSE OPEN
OFF NORM CLOSED
OFF NORM OPEN
OFF REVERSE CLOSED
OFF REVERSE OPEN
Test Conditions Patient Lead or Combination Measured
System Power Grounding/PE
Right Arm
(RA)
Left Arm
(LA)
Left Leg
(LL)
RA+LA+LL
Mains on open
RF on open
Table 10-16 Testing Power Conditions
ECG Power Meter’s Polarity Switch Meter’s Neutral Switch