1 - Introduction
16 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
1.5.2
Test Cartridges
GEM Hemochron 100 test cartridges are single‑use disposable test devices with a sample well for
application of samples.
NOTE: Test cartridges are not supplied with the system. They must be ordered separately. See “Materials
Required, but Not Provided” (page 19).
During a patient or LQC test, the instrument prompts the
Operator
to insert a test cartridge into the
instrument. After the instrument warms the test cartridge, it prompts the
Operator
to apply the sample into the
sample well of the test cartridge. For a description of test cartridge components, see
Figure 3
and
Table 2
.
Figure 3
: Test Cartridge Components
4. Sample Well
5. Overow Area
3. Waste Channel
6. Test Channel
1. Barcode Label
2. Detection Windows
Table 2: Test Cartridge Components
No. Component Description
1 Barcode Label The barcode label contains the test cartridge lot number, expiration date, and test
type. When a test cartridge is inserted into the test cartridge insertion chamber, the
instrument reads this information. Information on the label is also human-readable.
2 Detection Windows An internal camera analyzes the sample through the detection windows.
3 Waste Channel The instrument vacuum pump draws excess sample into the waste channel. This
ensures that the sample volume is correct.
4 Sample Well The location where the sample is applied.
5 Overow Area Excess sample ows into this area and then into the waste channel.
6 Test Channel The instrument vacuum pump draws the sample into the test channel where it is
mixed with reagent.
There are 2 types of test cartridges are available. The type of test cartridge must match the type of test that
will be performed:
ACT+ (Catalog #: 000GACT+)
ACT‑LR (Catalog #: 000GACT‑LR)
NOTE: Refer to the package insert accompanying the GEM Hemochron 100 test cartridges for storage,
handling, and disposal instructions.
1.5.3
Centralized Conguration Manager(CCM) Software
The software titled
Centralized Conguration Manager for GEM Hemochron 100
is intended to be used
with a PC (Microsoft Windows
®
operating system). Although the instrument menus can be used to
congure some of the instrument settings, CCM can be used to congure more advanced settings and
send those settings over the network to 1 or more instruments at the same time. For information, see
“Centralized Conguration Manager (CCM) Software” (page 71)
.
NOTE: This software is not included in the instrument box. It must be ordered separately. See “Optional
Materials” (page 19).
1.5.4
Liquid Quality Control (LQC)Product for an LQCTest
Specic LQC product is required when an LQC test is run. For information about the LQC product, see
“LQC Products” (page 53)
.
NOTE: LQC product is not supplied with the system. It must be ordered separately. See “Materials
Required, but Not Provided” (page 19).