1 - Introduction
16 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
1.5.2
Test Cartridges
GEM Hemochron 100 test cartridges are single‑use disposable test devices with a sample well for
application of samples.
NOTE: Test cartridges are not supplied with the system. They must be ordered separately. See “Materials
Required, but Not Provided” (page 19).
During a patient or LQC test, the instrument prompts the
Operator
to insert a test cartridge into the
instrument. After the instrument warms the test cartridge, it prompts the
Operator
to apply the sample into the
sample well of the test cartridge. For a description of test cartridge components, see
Figure 3
and
Table 2
.
Figure 3
: Test Cartridge Components
4. Sample Well
5. Overow Area
3. Waste Channel
6. Test Channel
1. Barcode Label
2. Detection Windows
Table 2: Test Cartridge Components
No. Component Description
1 Barcode Label The barcode label contains the test cartridge lot number, expiration date, and test
type. When a test cartridge is inserted into the test cartridge insertion chamber, the
instrument reads this information. Information on the label is also human-readable.
2 Detection Windows An internal camera analyzes the sample through the detection windows.
3 Waste Channel The instrument vacuum pump draws excess sample into the waste channel. This
ensures that the sample volume is correct.
4 Sample Well The location where the sample is applied.
5 Overow Area Excess sample ows into this area and then into the waste channel.
6 Test Channel The instrument vacuum pump draws the sample into the test channel where it is
mixed with reagent.
There are 2 types of test cartridges are available. The type of test cartridge must match the type of test that
will be performed:
● ACT+ (Catalog #: 000GACT+)
● ACT‑LR (Catalog #: 000GACT‑LR)
NOTE: Refer to the package insert accompanying the GEM Hemochron 100 test cartridges for storage,
handling, and disposal instructions.
1.5.3
Centralized Conguration Manager(CCM) Software
The software titled
Centralized Conguration Manager for GEM Hemochron 100
is intended to be used
with a PC (Microsoft Windows
®
operating system). Although the instrument menus can be used to
congure some of the instrument settings, CCM can be used to congure more advanced settings and
send those settings over the network to 1 or more instruments at the same time. For information, see
“Centralized Conguration Manager (CCM) Software” (page 71)
.
NOTE: This software is not included in the instrument box. It must be ordered separately. See “Optional
Materials” (page 19).
1.5.4
Liquid Quality Control (LQC)Product for an LQCTest
Specic LQC product is required when an LQC test is run. For information about the LQC product, see
“LQC Products” (page 53)
.
NOTE: LQC product is not supplied with the system. It must be ordered separately. See “Materials
Required, but Not Provided” (page 19).