3 - Instrument Settings
40 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
41 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
3.3.14
PatientID(PID) Settings
Supervisors
can use the instrument menus to require
Operators
enter PID to run tests.
When
Operators
are required to scan the PID, the user is prompted to activate the barcode scanner. For
more information about using the barcode scanner, see
“Using the Barcode Scanner” (page 24)
.
In addition to using the instrument menus to congure PID settings,
Supervisors
can also use CCM to
congure these same settings and to congure other more advanced PID settings. See
“Centralized
Conguration Manager (CCM) Software” (page 71)
.
3.3.14.1
Require Operators to Enter a PID and Specify the Method
Supervisors
can require
Operators
to enter a PID before they run a patient test. By choosing whether
or not to require
Operators
to enter a PID, a
Supervisor
also species the method the
Operator
is
allowed to use to enter the PID.
1 Select (Settings) > Supervisor Settings > PID Entry Preference.
2 In the dialog box (
Figure 37
), select the appropriate option:
● Not Required – Before
Operators
run a patient test, they are NOT required to enter a PID.
● Required (Scan or Manual Entry) – Before
Operators
run a patient test, they are required to
enter a PID. They can scan a compatible barcode or manually enter the PID.
● Scan Only – Before
Operators
run a patient test, they are required to enter a PID by scanning
a compatible barcode.
Figure 37
: PID Entry Preference
3.3.14.2
Reuse a PID (Automatic Entry of the Most Recent PID)
If a
Supervisor
requires an
Operator
to enter a PID when they run a patient test, they can choose to
have the instrument automatically enter the PID that was entered for the previous test. This is a
convenience that keeps the
Operator
from having to manually reenter the PID when they run a series
of tests. Instead, the
Operator
only needs to review the PID that was automatically entered and then
accept it.
PID reuse can remain enabled for no less than 1 hour and no more than 240 hours. The factory default
setting duration is 0 (zero) hours. Meaning, PID reuse will not go into eect until a
Supervisor
species
a duration other than 0 (zero). See
“Specify PID Reuse Duration” (page 42)
.
NOTE: If the instrument automatically shuts down or if an Operator manually shuts it down, the
PID reuse duration is automatically reset to 0 (zero).
Allow Reuse of a PID:
1 Select (Settings) > Supervisor Settings > PID Reuse (During Patient Test).
2 Select On to enable PID reuse:
● O (Toggle O) – Disabled.
● On (Toggle On) – Enabled (the PID will automatically be entered into the dialog box).