3 - Instrument Settings
42 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
Specify PIDReuse Duration:
1 Select (Settings) > Supervisor Settings > PID Reuse Duration.
2 In the dialog box (
Figure 38
), scroll the Hours and Minutes to select the duration. The minimum
duration is 1 hour; maximum is 240 hours.
Figure 38
: PID Reuse Duration
3 Select OK.
NOTE: After a period of instrument inactivity, the PID reuse duration is automatically reset to 0 (zero).
NOTE: PID reuse will not go into eect until a Supervisor species the period of time the feature needs
to remain enabled.
3.3.15
Clear Test Results
A maximum of 10,000 test results (patient, LQC, and EQC tests) can be stored in the instrument database.
Supervisors
can clear (delete) test results.
CAUTION: Deleted test results cannot be recovered.
If the database is full and an
Operator
runs a test, the oldest test result is overwritten. Periodically clear the
database by doing the following:
● (Optional) Back up the test results that need to be deleted by sending the test results to the network,
see
“Database” (page 63)
.
● Enable the setting to erase (delete) the Database in CCM, see
Table 14 (page 83)
.
● Clear the test results.
To erase (delete) Database test results (patient, LQC and EQC tests), 2 steps are required:
1 Enable the setting to erase (delete) the Database in CCM, see
Table 14 (page 83)
.
2 Follow instructions below on the instrument:
Clear All Test Results:
1 Select (Settings) > Supervisor Settings > Clear All Test Results.
2 In the dialog box, select Delete to delete all test results.
Clear All Patient Test Results:
1 Select (Settings) > Supervisor Settings > Clear All Patient Test Results.
2 In the dialog box, select Delete to delete all patient test results.