3 - Instrument Settings
40 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
3.3.13.2
Reuse an OID (Automatic Entry of the Most Recent OID)
If a
Supervisor
requires
Operators
to enter their OID when they run a test, they can choose to have the
instrument automatically enter the OID that was entered for the previous test. This is a convenience
that keeps the
Operator
from having to manually reenter their OID when they run a series of tests.
Instead, the
Operator
only needs to review the OID that was automatically entered and then accept it.
OID reuse can remain enabled for no less than 30 minutes and no more than 720 hours. The factory
default setting duration is 0 (zero) hours. Meaning, OID reuse will not go into eect until the
Supervisor
species a duration other than 0 (zero). See
“Allow OID Reuse During Patient Testing” (page 40)
.
NOTE: If the instrument automatically shuts down or if an Operator manually shuts it down, the OID
reuse duration is automatically reset to 0 (zero).
Allow OIDReuse During Patient Testing:
1 Select (Settings) > Supervisor Settings > OID Reuse (During Patient Tests).
2 Select the appropriate option to enable or disable OID reuse:
● O (Toggle O) – Disabled.
● On (Toggle On) – Enabled (the OID will automatically be entered into the dialog box).
Allow OIDReuse During LQCTesting:
1 Select (Settings) > Supervisor Settings > OID Reuse (During LQC Test).
2 Select the appropriate option to enable or disable OID reuse:
● O (Toggle O) – Disabled.
● On (Toggle On) – Enabled (the OID will automatically be entered into the dialog box).
Specify OIDReuse Duration:
1 Select (Settings) > Supervisor Settings > OID Reuse Duration.
2 In the dialog box (
Figure 36
), scroll the Hours and Minutes to select the duration. The minimum
duration is 30 minutes; maximum duration is 720 hours.
Figure 36
: OID Reuse Duration
3 Select OK.
NOTE: After a period of instrument inactivity, the OID reuse duration is automatically reset to 0 (zero).
NOTE: OID reuse will not go into eect until the Supervisor species the period of time the feature is to
remain enabled.