5 - Quality Control(QC)
56 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
57 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
5.4
Run an EQCTest
Before Running an EQCTest:
NOTE: If a test cartridge was left in the instrument from a previous test, remove and dispose in a
biohazardous waste container.
Run an EQCTest:
1 On the Home screen, select (Quality Control).
2 Select the EQC tab (
Figure 58
).
Figure 58
: Select EQC Tab
EQC Tab
3 Select (Start Test).
The EQC test starts. Information about test progress is displayed (
Figure 60
).
If the EQC tab displays (EQC Interval Expired) or (Testing Failed), the EQC Interval has
expired or the previous EQC test failed (
Figure 59
).
Figure 59
: LQC and Patient Testing is Not Allowed
NOTE: The QC status is also displayed on the QC icon located on the Home screen. (QC Expired) for
Expired and (QC Failed) for Failed.
NOTE: To abort the test, select (Abort Test) at any time (Figure 60).
Figure 60
: EQC Test in Progress
Select this icon
to abort the test
at any time.
NOTE: By pressing (Home) or (Back), the test can be canceled.