5 - Quality Control(QC)
56 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
To Clear an EQCLockout:
1 Select the EQC Settings tab (
Figure 57
).
Figure 57
: Clear an EQC Lockout in CCM
2 Select Yes in the Clear EQC Failure Count drop-down list.
3 Select Yes in the Clear Number of Patient Tests Used After EQC Expiration drop-down list.
4 Select OK.