3 - Instrument Settings
32 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
33 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
Manually Set the Time:
1 Select (Settings) > Supervisor Settings > Instrument Settings > Date and Time.
2 Set the toggle for Automatic Date & Time to O (Toggle O).
3 Select Set Time.
4 In the dialog box (
Figure 25
), scroll the values for the hour and minute.
Figure 25
: Set Time
5 Select OK.
Set the Format for the Date and Time:
● To set the date format (MM/DD/YYYY, DD/MM/YYYY, or YYYY/MM/DD), use CCM. See
“Congure
Instruments (Instrument Settings)” (page 83)
.
● To set the time format (12 hour or 24 hour), use CCM.
Set the Time Zone:
● To set the time zone, use CCM.
3.3.6
View Battery Usage Information Based on System Resource
The instrument records the percentage of battery power used based on the resource that is used. An
example of a resource that uses battery power is the backlight system that illuminates the display screen.
● To view a list of resources that use battery power, select (Settings) > Supervisor Settings >
Instrument Settings > Battery.
3.3.7
Connectivity Settings
Supervisors
can congure the instrument to be connected to a network or other device. See
Table 7
.
Table 7: Connectivity Options
Connection Type Description
Ethernet
(Wired Network)
Use an Ethernet cable to connect the instrument to a Local Area Network (LAN). See
“Connect to a Wired Network (Ethernet)” (page 34)
.
Wi-Fi
(Wireless Network)
Use Wi-Fi to connect the instrument to a Wireless Local Area Network (WLAN) through
802.11 b/g/n connections at 2.4 & 5 GHz. See
“Connect to a Wireless Network (Wi-Fi)”
(page 35)
.
POCT1-A Using Ethernet or Wi‑Fi,
Supervisors
can connect the instrument to a POCT1-A-compliant
network for use with middleware. See