8 - Centralized Conguration Manager(CCM) Software
98 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
8.4.5
Update Instrument Software
Use CCM to update the instrument software (rmware).
1 Get the build package that contains the latest instrument software from the manufacturer:
GEMH100.zip
2 Log into CCM as a
Supervisor
.
3 Ensure the instrument that needs to be updated is connected to the network and the power supply
cable.
NOTE: If (Ethernet) or (Wi-Fi) is displayed in the status bar of the instrument screen, the
instrument is connected to the network.
4 Under Instrument Selection and Control (
Figure 82 (page 78)
), in the drop-down list, select the
instrument that needs to be updated.
5 The icon to the right of the instrument name indicates whether a network connection is established
between the instrument and CCM.
Network connection is established. Go to the next step.
Network connection is not established. See
“Install CCM Software and Congure Firewalls” (page 73)
.
6 Click the Update Software button.
7 Select the instrument software build package:
GEMH100.zip
8 Click Open and then select OK in the dialog box to conrm the download.
9 The updating process begins. During the process, the instrument is automatically shut down and
restarted twice, so allow time for the process to run.
8.4.6
Update Wi‑Fi Enterprise Certicate
Use CCM to Update Wi‑Fi Enterprise Certicate:
1 Log into CCM as a
Supervisor
.
2 Ensure each instrument that needs to be updated is connected to the network.
NOTE: If (Ethernet) or (Wi-Fi) is displayed in the status bar of the instrument screen, the
instrument is connected to the network.
3 Under Instrument Selection and Control (
Figure 82 (page 78)
), in the drop-down list, select the
instrument, group of instruments, or all instruments that need to be updated with the Wi‑Fi Enterprise
Certicate.
4 The icon to the right of the instrument name indicates whether a network connection is established
between the instrument and CCM.
Network connection is established. Go to the next step.
Network connection is not established. See
“Install CCM Software and Congure Firewalls” (page 73)
.
5 Click the Update Certicate button.
6 Select the Wi‑Fi Enterprise Certicate le, click Open, and then select OK in the Update Certicate
dialog box (
Figure 99
) to upload the certicate onto the instrument.
Figure 99
: CCM – Update Certicate