8 -  Centralized Conguration Manager(CCM) Software
82 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
3  Enter the following test cartridge lot information:
 ● In the Test Type drop-down list, select the appropriate test type for the test cartridge 
lot (ACT+ or ACT-LR).
 ● Enter the Cartridge Lot Number.
 ● Enter the Expiration Date.
 ● Click OK. The test cartridge lot is added to the list of Cartridge Lots.
4  Click OK to save the changes (
Figure 86 (page 81)
).
NOTE: To edit or delete a test cartridge lot, select the test cartridge lot from the list (Figure 86 
(page 81)) and the appropriate button (Edit or Delete).
8.4.2.1.2 
Congure LQCLots Data
1  Click the LQC Lots button.  
The LQC Lots dialog box (
Figure 88
) is displayed.
Figure 88
: CCM – LQC Lots Dialog Box
2  Click New.
3  Enter the following LQC lot information:
 ● In the Test Type drop-down list, select the appropriate test type for the LQC lot (ACT+ or 
ACT-LR).
 ● In the LQC Level drop-down list, select the appropriate level for the LQC lot (Level 1 
[Normal] or Level 2 [Abnormal]).
 ● Enter the LQC Lot Number.
 ● Enter the Ranges (Low and High).
 ● Enter the Expiration Date.
 ● Click OK and the LQC lot is added to the list of LQC Lots.
4  Click OK to save the changes.
NOTE: To edit or delete an LQC lot, select the LQC lot from the list (Figure 88) and the appropriate 
button (Edit or Delete).
NOTE: When editing an LQC lot, if the test type is changed, the Range Low and Range High 
values will be reset to 0 (zero). Please ensure the appropriate values are entered in this text eld 
before saving.