8 - Centralized Conguration Manager(CCM) Software
71 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
70 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
Δ 8
Centralized Conguration Manager(CCM) Software
8.1
CCM Introduction
Centralized Conguration Manager
is a PC‑based software that is used to congure the GEM Hemochron 100
instruments.
Although the instrument menus can be used to congure some of the instrument settings, CCM can be used
to congure more advanced settings and send the settings over the network to 1 or more instruments
at the same time.
CCM also allows
Supervisors
to do the following:
Generate a conguration data report – Use CCM to generate a report containing all of the conguration
data associated with 1 or more instruments. Reports can be saved as a comma‑separated value le (.csv)
and opened in a spreadsheet.
Reuse conguration dataAfter conguration data is created, use CCM to export to a le for reuse.
Update the instrument software – Use CCM to install a new version of instrument software (rmware).
When updating the software on the instruments, multiple instruments can be updated at the same time.
The CCM software application consists of 2 modules:
Administrative Settings – To create and manage
Admins
and
CCM Supervisor
accounts.
Conguration Settings – To congure instruments.
NOTE: Users log into each module separately. Users must log out of 1 module before logging into the
other module.
8.1.1
GEM Hemochron 100 System User Accounts
There are 4 types of accounts are associated with the GEM Hemochron 100 system. 2 accounts are for the
CCM software application and 2 accounts are for instrument(s):
CCM Accounts:
CCM Administrator
CCM Supervisor
Instrument Accounts:
Instrument
Operator
Instrument
Supervisor
NOTE: The Supervisor can use the instrument settings menu to congure that specic instrument or use
CCM to congure more advanced instrument settings and send those settings over the network to 1 or
more instruments at the same time.
CCM Administrator (Admin):
Admins
use the Administrative Settings module (
Figure 75 (page 74)
) to do the following:
Create, edit, and delete accounts for other
Admins
.
Create, edit, and delete accounts for
CCM Supervisors
.
A factory default setting
Admin
account exists in CCM:
Username
admin
Password
000
The rst user to log into the software must log in using the factory default setting
Admin
account
(
admin/000
). Once logged in, the user will be forced to change the password. The new password should be
recorded and stored in a safe place. The new
Admin
is able to create other
Admin
and
CCM Supervisor
accounts.
Admin
account information is stored in an encrypted le on the same computer that installed the
CCM software. When the software is upgraded, this le remains on the system so that these accounts
do not need to be recreated. Therefore, the
Admin
default password will not change unless it is changed
by the
Admin.