4 Operation and use
4.1 Notes on use
NOTE
The graphics in these instructions for use show the available functions of the
touchscreen user interfaces. Depending on the OR table being operated, there
may be differences between the user interface described here and the actual user
interface.
On every occasion, before adjusting the OR table, check the set patient orientation
[8Page10] and adapt as necessary.
As a general principle, do not perform adjustments with two control devices at the same time.
In the event that multiple control devices are accidentally used simultaneously, the override
control panel and corded controls take priority over the infrared devices. If equally ranked
control devices are used at the same time, adjustment may be interrupted or stopped.
The adjustment speed may be modified according to the overall load and protrusion.
In the event of large overall loads/protrusion, the speed of lowering actions is increased and
the speed of raising actions is reduced.
The adjustment functions are executed for as long as the appropriate adjustment button is
pressed. If the adjustment button is released, the adjustment is stopped.
The capacity of a fully charged rechargeable battery is sufficient for several days of OR oper-
ation or around 2.5 hours of continuous operation.
In order to avoid empty batteries of the control device during surgery, we recommend storing
the operating device in the charging cradle when not in use.
When the battery charge is ≤20 %, the touchscreen shows the message Remote control
battery low for approximately 5 seconds. The control device remains switched on and can
be used again.
When the battery charge is ≤5 %, the touchscreen shows the message Remote control
battery empty. The control device switches itself off after 5 seconds. After 10 seconds in the
charger cradle, the control device can be switched back on and you can continue to use it in
the charger cradle.
4.2 Transmission power
The transmission signal of the control device is sufficient to adjust the connected OR table from a
greater distance. in wireless operation.
If the button is pressed and the desired adjustment of the OR table is not carried out and the
status symbol for signal exchange between the OR table and control device [8Page29] is not
displayed, the infrared receiver on the column may be covered (for instance by a person). In this
case, change the direction or position of the control device slightly and repeat the adjustment.