EasyManua.ls Logo

Getinge 1009.25B0 - Menu [Drive]; Menu [Active Cable Charging]; Safety Settings] Menu; Activate; Deactivate Safety Functions

Getinge 1009.25B0
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
Operating menu
Main menu [Menu]
46 / 92
1009.25B0
IFU 1009.25 EN 01
2. Touch the OK button.
Ø The function is activated.
Ø External devices cannot activate the
operating table.
Unlock activation via external devices
Prerequisites:
The Lock OR table menu is open.
The function is activated.
1. Touch the External Control Lock button.
Ø The status display is greyed out.
2. Touch the OK button.
Ø The function is deactivated.
Ø External devices can activate the oper-
ating table.
5.3.6 Menu [Drive]
NOTIC E
The menu Drive can either be opened via the main menu or, if the operating table
is already in UNLOCK, by pressing the [UNLOCK] button.
For further information, see section Moving the OR table using autodrive [UNLOCK]
[8Page70].
5.3.7 Menu [Active Cable Charging]
For further information, see section Charging the control device via the cable [8Page21].
5.3.8 [Safety settings] menu
5.3.8.1 Activate / deactivate safety functions
NOTIC E
Only deactivate the safety functions in exceptional cases. Operate the OR table
with the safety functions turned off at your own risk.
NOTIC E
When a safety function is deactivated, a warning message will be displayed con-
tinuously on the start screen of the control device.
NOTIC E
The OR table goes into standby mode approx. 15 seconds after the last time a but-
ton is pressed. Once the OR table is reactivated, the safety function is activated
once again as well.

Table of Contents