Maintenance
Malfunctions and troubleshooting
80 / 92
1009.25B0
IFU 1009.25 EN 01
No. Inspection Defects are present No defects
6 Are there damages to the insu-
lation of electrical cables?
1. Do not continue to use the
product.
2. Notify Getinge authorised
service.
Comment:
7 (Space for other tests)
1.
Comment:
Tab.11: Visual and functional inspections
8.2 Malfunctions and troubleshooting
8.2.1 Control device
NOTE
Software version [8Page40] and error messages [8Page39] can be called up
in the control device.
No
.
Appearance / Inspection Activity / Rectification / Measure
1 Weak transmission power IR window dirty.
1. Clean the IR window of the control device.
IR window scratched / damaged.
1. Have the control device checked by an author-
ised service technician [8Page84].
2 No transmission signal during use as
an IR control device
No IR transmission
1. Use the control device with cable [8Page21].
2. Have the control device checked by an author-
ised service technician [8Page84].
3 Short battery run time / frequent
charging required
Defective batteries
1. Have the battery replaced by an authorised ser-
vice technician or in-house technician
[8Page84]. Request and follow service in-
structions.
Tab.12: Faults and troubleshooting