Maintenance
Malfunctions and troubleshooting
8
1009.25B0
IFU 1009.25 EN 01
81 / 92
No
.
Appearance / Inspection Activity / Rectification / Measure
4 Control device does not charge, or
charges slowly (≥6hours to charge
level 100%)
Control device is not in the charging cradle (mobile
charging station)
1. Insert the control device into the charger so that
the green LED flashes.
Holding bracket of the control device is bent (sta-
tionary charging station)
1. Have the holding bracket replaced by an author-
ised service technician or in-house technician
[8Page84]. Request and follow service in-
structions.
5 Selected adjustment functions are
not executed
Reception disturbed.
1. Change position and press the buttons again.
Incorrect connection with OR table.
1. Check IR code [8Page35].
2. Re-connect the control device with the OR table
[8Page31].
3. Restart the control device [8Page63].
The control device is defective.
1. Have the control device checked by authorised
service technicians.
6 Control device no longer respond-
ing / touchscreen is black
System failure of the control device.
1. Restart the control device [8Page63].
- or -
2. operate the OR table using the override control
panel or another control device.
The control device is defective.
1. Have the control device checked by authorised
service technicians.
Tab.12: Faults and troubleshooting