9
Technical specifications
Electromagnetic compatibility (EMC)
86 / 92
1009.25B0
IFU 1009.25 EN 01
9.5 Electromagnetic compatibility (EMC)
The product is designed for operation in the environments specified in the intended use. The user
must ensure that the product is operated in such an environment.
9.5.1 Electro magnetic transmission and resistance to interference
Standard Compliance
Medical electrical equipment – Part 1-2:
General requirements for basic safety, in-
cluding significant performance features –
collateral standard: Electromagnetic dis-
turbances – requirements and testing
IEC 60601-1-2
Conformity without deviations
Interference voltages at the power supply
connection (conducted emissions), elec-
tromagnetic interference (radiated emis-
sions)
CISPR 11
Group 1
Class A
9.5.2 Resistance to electro magnetic interference - Testing process
Test / Standard Test level
Interference immunity to static discharge
IEC 61000-4-2
±8kV contact discharge
±2kV, ±4kV, ±8kV, ±15kV air discharge
Interference immunity to high-frequency
electromagnetic fields
IEC 61000-4-3
3V/m
80MHz – 2.7GHz
80% AM at 1kHz