EasyManua.ls Logo

HBM QuantumX MX840A - Frequency Measurement Without Directional Signal (Single−Pole Mode)

HBM QuantumX MX840A
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
135Connection
HBMQuantumX
6.11.21 Frequency measurement without directional signal
(singlepole mode)
Supported by modules: MX840, MX840A, MX440A, MX460
9
4
Hsg.
Cable color code:
wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
10
Ground
Connector 1
Md
Cable shield
f
1
(+)
5
1
4
rd
wh
gy
5
6
5 V
3.5 V
1.5 V
Thresholds
Thresholds
f
1
Schematic diagram; Singlepole mode
0 V
Hsg. = housing
HBM torque transducer
6
1
123
1-wire EEPROM
(optional)
view from below
1
5
6
10
11
15
2 Data
3 No function
1
Adjustable sensor supply: Pin 12: 5 V ... 24 V
Pin 11: 0 V