5 June 2020 CasTemp Wireless including CasTemp Superheat Page 45
Table 1: List of available fields in the telegram composer
The ID of the QUBE CTW module that
is paired during the measurement
Equals 0 if no error is active, Equals 8
if an error is active
Bit 1 = CTW is open circuit
Bit 2 = CTW cold junction temperature
> 85 °C
Bit 3 = CTW lost transmission
Bit 4 = CTW battery charge is critical
(charge < 10%)
Bit 5 = CTW signal strength is critical
(strength < 40dB)
Bit 6 = Reserved for future use
Bit 7 = Instrument is not paired with
CTW
Bit 1 = CTW cold junction temperature
not in range (-20°C <= x <= 60°C)
Bit 2 = Reserved for future use
Bit 3 = Reserved for future use
Bit 4 = Reserved for future use
Bit 5 = Reserved for future use
Bit 6 = Reserved for future use
Bit 7 = Reserved for future use
Bit 8 = Reserved for future use
The remaining charge of the battery of
the QUBE CTW module in %
The timestamp of the measurement as
it is recorded by the instrument
Text, hh:mm:ss tt, 11
bytes
The datetimestamp of the
measurement as it is recorded by the
instrument
Text, dd/mm/yyyy
hh:mm:ss, 19 bytes
Cold junction temperature
The temperature measured by the
QUBE CTW
Text format, xxxx.y, 6
bytes
The voltage of the battery of the QUBE
CTW
The name of the instrument as it is
configured in the settings
Text, 1 byte per
character