5 June 2020 CasTemp Wireless including CasTemp Superheat Page 45
Table 1: List of available fields in the telegram composer
Field Name
Explanation
Format and Size
Units
ModuleID
The ID of the QUBE CTW module that
is paired during the measurement
Text, 4 bytes []
ErrorState
Equals 0 if no error is active, Equals 8
if an error is active
1 byte
ErrorPresent[0]
Bit 1 = CTW is open circuit
1 byte
Bit 2 = CTW cold junction temperature
> 85 °C
1 byte
Bit 3 = CTW lost transmission
1 byte
Bit 4 = CTW battery charge is critical
(charge < 10%)
1 byte
Bit 5 = CTW signal strength is critical
(strength < 40dB)
1 byte
Bit 6 = Reserved for future use
1 byte
Bit 7 = Instrument is not paired with
CTW
1 byte
Bit 8 = Not used
1 byte
ErrorPresent[1]
Bit 1 = CTW cold junction temperature
not in range (-20°C <= x <= 60°C)
1 byte
Bit 2 = Reserved for future use
Bit 3 = Reserved for future use
Bit 4 = Reserved for future use
Bit 5 = Reserved for future use
Bit 6 = Reserved for future use
Bit 7 = Reserved for future use
Bit 8 = Reserved for future use
ErrorPresent[2]
Not currently used
1 byte
ErrorPresent[3]
Not currently used
1 byte
BatteryCharge
The remaining charge of the battery of
the QUBE CTW module in %
Text, xyz% 4 bytes
%
SystemTimestamp
The timestamp of the measurement as
it is recorded by the instrument
Text, hh:mm:ss tt, 11
bytes
(tt = Am
or PM)
SystemDateTimestamp
The datetimestamp of the
measurement as it is recorded by the
instrument
Text, dd/mm/yyyy
hh:mm:ss, 19 bytes
CJTemp
Cold junction temperature
Text, xxx.y , 5 bytes
[degC] or
[degF]
Temperature
The temperature measured by the
QUBE CTW
Text format, xxxx.y, 6
bytes
[degC] or
[degF]
BatteryVoltage
The voltage of the battery of the QUBE
CTW
Text, xxxx, 4 bytes
[mV]
InstrumentName
The name of the instrument as it is
configured in the settings
Text, 1 byte per
character
SignalStrength
Signal strength of CTW
Text, xyz 3 bytes

Table of Contents