5 June 2020 CasTemp Wireless including CasTemp Superheat Page 9
8 Technical Data
8.1 Operating Specification
QUBE CTW module
Minimum Temperature
Maximum Temperature
Specified Operating
Temperature Range
-20
0
C
60
0
C
Do not exceed Temperature
Range
-30
0
C
85
0
C
Battery Charging
0
0
C
60
0
C
The sensor contact block has a maximum operating temperature of 200
0
C
The wireless transmission system will not operate at temperatures above 85
0
C
The Qube CTW module must NOT be exposed to temperatures above 85
0
C to avoid risk
of permanent damage to electronics and batteries
Battery
LiFePO
4
3.2V 3Ah
Charging
5V dc 850mA via USB socket
Wireless module
2.4 GHz (2.4-2.483 GHz) frequency hopping
8.2 Measurement Specification
Measurement application
Continuous temperature
measurement in tundishes
Measurement input
Thermocouple type B
CasTemp sensor using QUBE
CTW module
Temperature input range
0°C to 1810°C
Measurement accuracy
Temperature +- 1°C
For ambient temperature
between 0 and 50°C
Data output
TCP, RS232, Profibus, EIP
Housing, dimensions and
weight
Metal housing protection IP65
Operating data
Power supply 110V/220V
50/60 Hz