4.12 Maintaining Measurement Precision (Self-Calibration)
93
4
When self-calibration operation is set to manual, be sure to perform self-calibration if the
temperature of the environment in which the instrument is operating changes by 2 degrees
or more. (Accuracy cannot be guaranteed if self-calibration is not performed.)
Even if the temperature variation in the operating environment is less than 2 degrees, self-
calibration should be performed at a 30-minute interval.
2
Open the System Setting Screen.
3
Configure self-calibration operation.
4
Return to the Measurement screen.
Move the cursor to the [SYS]
tab with the left and right cur-
sor keys.
2
Configure automatically.
(default)
Configure manually.
Selection
1
Return to the
Measurement screen.