10.2 Timing Chart
192
Auto setting operation
Self-calibration starts immediately after measurement completes and is finished in 5 ms.
One TRIG signal received during self-calibration is held, and measurement will start after
the self-calibration completes.
During auto-scan operation, self-calibration starts only after scanning completes. Self-cal-
ibration will not be performed after each channel is measured.
A 400 ms self-calibration is performed immediately after switching from MANUAL to
AUTO. Do not input the TRIG signal during that interval.
If there is at least 5 ms of extra time in the measurement interval
If the TRIG signal is received during self-calibration
Measurement
processing
Measure-
ment
Self-
calibration
Self-
calibration
???
Measure-
ment
Self-
calibration
Self-
calibration
Measure-
ment
Measure-
ment
Measurement
processing
The TRIG signal received during self-cali-
bration is held.