EasyManua.ls Logo

Hioki RM3545 - Page 351

Hioki RM3545
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix 19 Calibration
A39
Appendix
D/A output
Calibration points
*1 Without 0ADJ: Only with OVC off
Equipment Calibration point Manufacturer Standard model
Voltmeter
0 : 0 V
HIOKI Equivalent to 3237
1 : 1 V
Output cable
Wiring resistance: 500 m or less
(circuit resistance)
Range Calibration point OVC
Measure
ment
Current
100 M
range high-
precision
mode
0ADJ
Resistance
measurement
(Low Power: OFF)
10 m 0 , 10 m ON, OFF - -
With or
without
*1
100 m 0 , 100 m ON, OFF High, Low -
With or
without
*1
1 0 , 1 ON, OFF High, Low -
With or
without
*1
10 0 , 10 ON, OFF High, Low -
With or
without
*1
100 0 , 100 ON, OFF High, Low -
With or
without
*1
1000 0 , 1 k ON, OFF - -
With or
without
*1
10 k 0 , 10 k OFF - - -
100 k 0 , 100 k OFF - - -
1000 k 0 , 1 M OFF - - -
10 M 0 , 10 M OFF - - -
100 M 0 , 100 M OFF - ON, OFF -
1000 M 0 , 1000 M OFF - OFF -
Resistance
measurement
(Low Power: ON)
1000 m 0 , 1 ON - - -
10 0 , 10 ON - - -
100 0 , 100 ON - - -
1000 0 , 1 k ON - - -
Temperature
(thermistor)
25°C: 2186.0
input
Temperature
(analog input)
10°C: 0.1 V input
100°C: 1 V input
D/A output 1
0 : 0 V input
1 : 1 V input

Table of Contents

Other manuals for Hioki RM3545

Related product manuals