8.7 Command Message Description
192
Example
Command
:CONFIGURE:MTIME 5
Sets the measuring time of automatic measurement to 5 seconds.
Query
Response
:CONFigure:MTIMe?
:CONFIGURE:MTIME 5 (Headers: ON)
5 (Headers: OFF)
The measuring time of automatic measurement has been set to 5
seconds.
Error
An execution error occurs in the following cases:
When the measurement method is set to "manual measurement"
When the measurement mode is not selected
When a mode other than "leakage current mode" is set
In automatic measurement(Command)
Note
Use this command after executing the :MODE command to activate the mea-
surement mode.
The unit is [s]. Fractions are rounded up when 0.5 or greater and rounded off
when less than 0.5.
Sets and queries the measuring time of automatic measurement
Other manual measurements: Sets and queries the application of 110% of
rated voltage(only ST5540)
Syntax
Command
Query
Response
:CONFigure:OTHer <characters>
:CONFigure:OTHer?
<characters>
<characters> = OFF/NAPPly/RAPPly
OFF : Application of 110% voltage : none
NAPPly : Application of 110% voltage : Positive phase
RAPPly : Application of 110% voltage : Negative phase
Function
Command Other manual measurements: Sets application of 110% of rated voltage.
Query Other manual measurements: Returns application of 110% of rated voltage.
Example
Command
:CONFigure:AUTO OFF;:CONFigure:OTHer NAPPly
Other manual measurements: Sets application of 110% of rated voltage to pos-
itive phase.
Query
Response
:CONFigure:AUTO OFF;:CONFigure:OTHer?
:CONFIGURE:OTHER NAPPLY (Headers: ON)
NAPPLY (Headers: OFF)
Other manual measurements : Application of 110% of rated voltage has been
set to positive phase.
Error
An execution error occurs in the following cases:
When the measurement method is set to "automatic measurement"
When a mode other than "leakage current mode" is set
When a Network other than "B1" or "B2" is set
When the measurement mode is not set to Touch current : Enclosure - Earth,
Touch current:Enclosure - Enclosure, Patient leakage current:SIP/SOP, Total
patient leakage current : SIP/SOP, Free current