8.7 Command Message Description
218
Syntax
Response <numerical value 6> = Target current(NR1)
0 : AC+DC
1 : AC
2 : DC
3 : ACPeak
<numerical value 7> =Other 110% voltage application
0 : Application of 110% voltage : none
1 : Application of 110% voltage : Positive phase
2 : Application of 110% voltage : Negative phase
(0 for ST5541)
<numerical value 8> =
Specific 110% voltage application
0 : Application of 110% voltage : none
1 : Application of 110% voltage : Positive phase
2 : Application of 110% voltage : Negative phase
(0 for ST5541)
<numerical value 9> =Switch
0 : S10=OFF, S12=OFF, S13=OFF
1 : S10=ON, S12=OFF, S13=OFF
2 : S10=OFF, S12=ON, S13=OFF
3 : S10=ON, S12=ON, S13=OFF
4 : S10=OFF, S12=OFF, S13=ON
5 : S10=ON, S12=OFF, S13=ON
6 : S10=OFF, S12=ON, S13=ON
7 : S10=ON, S12=ON, S13=ON
(0 for ST5541)
See "To Transfer All Saved Data to a Computer" (p. 243)
Function
Query Returns the saved data of the designated measurement mode for the specified
data unit.
The data is arranged as follows:
<1.Maximum value>, <2.Judgement>, <3.Power supply polarity>, <4.Equip-
ment under test status>, <5.Measurement network filter>, <6.Target current>,
<7.Other 110% voltage application>, <8.
Specific 110% voltage application>,
<9.Switch>
(unit of (1) maximum value : A)
In all cases, data items (1) through (9) are repeated as many times as the num-
ber of data units.
Returns only "0" when there is no saved data.
Example
Query
Response
:MEMory:READ:MEASURE? 1,ENCLosure1
:MEMORY:READ:MEASURE +2.345E-03,0,0,0,1,0,0,0,0,
+2.362E-03,0,1,0,1,0,0,0,0,
+2.510E-03,0,0,2,1,0,0,0,0,
+2.610E-03,1,1,2,1,0,0,0,0,
+2.456E-03,0,0,1,1,0,0,0,0,
+2.459E-03,0,1,1,1,0,0,0,0
(Headers : OFF)
+2.345E-03,0,0,0,1,0,0,0,0,
+2.362E-03,0,1,0,1,0,0,0,0,
+2.510E-03,0,0,2,1,0,0,0,0,
+2.610E-03,1,1,2,1,0,0,0,0,
+2.456E-03,0,0,1,1,0,0,0,0,
+2.459E-03,0,1,1,1,0,0,0,0
(Headers : ON)
The example of response data is shown below:
Reads saved data