8.7 Command Message Description
220
Saves the maximum value
Syntax
Command
:MEMory:SAVE:MAXimum
Function
Command Saves the date, Equipment name, Control number, Grounding class, Applied
part, network, measurement mode, Maximum value, test condition setting at
maximum value acquisition, and judgement result of the present measurement
in memory.
*
When network B1 or B2 is selected
Example
Command
:MEMory:SAVE:MAXimum
Saves the maximum value.
Error
An execution error occurs in the following cases:
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
• In automatic measurement
• When there is not enough memory space
• When the equipment name and control number are the same as those of
saved data, but the network, grounding class, and applied part differ
• When there is no data
Note
Use this command after executing the :MODE command to activate the mea-
surement mode.