Appendix3 List of default settings
A24
Network C
After selecting network and leakage current mode
default setting
Leakage current mode item Class I Class II Internally
powered
Earth leakage current
measurement manual
––
polarity
positive
phase
equipment status normal
filter ON1
target current ACDC
range AUTO
allowable value :
upper limit : normal
3.500 mA
allowable value :
upper limit : fault
3.500 mA
allowable value : lower limit :
ON/OFF : normal
OFF
allowable value : lower limit :
ON/OFF : fault
OFF
Touch current (Enclosure - Earth)
measurement manual manual –
polarity
positive
phase
positive
phase
–
equipment status normal normal –
filter ON1_U2 ON1_U2 ON1_U2
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper limit : normal 3.500 mA 250.0
A 250.0 A
allowable value : upper limit : fault 3.500 mA 250.0
A–
allowable value : lower limit :
ON/OFF : normal
OFF OFF OFF
allowable value : lower limit :
ON/OFF : fault
OFF OFF –
Touch current
(Enclosure - Enclosure)
measurement manual manual –
polarity
positive
phase
positive
phase
–
equipment status normal normal –
filter ON1_U2 ON1_U2 ON1_U2
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper limit : normal 3.500 mA 250.0
A 250.0 A
allowable value : upper limit : fault 3.500 mA 250.0 A–
allowable value : lower limit :
ON/OFF : normal
OFF OFF OFF
allowable value : lower limit :
ON/OFF : fault
OFF OFF –
*
( ) is the value after DC switching